| Measurement
Method |
Mainstream
NDIR |
| Package
Size |
2.0"
X 0.75" X 1.0" = 1.5 cu. In. |
| Weight |
1
oz. |
| Power
Consumption |
1.5
W, 12VDC |
| Operating
Temperature |
25
to 45 degrees C |
| Storage
Temperature |
-40
to 60 degrees C |
| Humidity |
0
to 85% non-condensing |
| Measured
Gases |
CO2
and N2O |
| Zero
Drift CO2 |
<
1.25 mmHg / 24 hour (1.6% of full scale)(STSP) |
| Zero
Drift N2O |
<
4% / 24 hour (STSP) |
| Span
Stability CO2 |
<
2 mmHg / year (STSP) |
| Span
Stability NO2 |
<
4% / year (STSP) |
| Accuracy:
CO2 |
+/-
3 mmHg 0-60 mmHg |
| +/-
5% of reading 61-100 mmHg |
| Accuracy:
N2O |
+/-5%
FS |
| Warm
Up Time |
<
90 seconds to measure gases |
| <
30 minutes to full specification |
| Airway
Adapter / Sample Cell |
polyethylene
body with 15/22 mm ANSI standard tapers |
| proprietary
plastic film windows reduce condensation problems |
| Re-Calibration
Frequency |
one
year |
| Response
Time: CO2 |
100
mS. Max. (10-90%, 200 ml/min.) |
| Response
Time: N2O |
100
mS. Max. (10-90%, 200 ml/min.) |
| Range:
CO2 |
0
to 100 mmHg |
| Range:
N2O |
0
to 100% |
| Noise:
CO2 |
1.5
mmHg RMS |
| Noise:
N2O |
15
mmHg RMS |